strengthening the materials core of manufacturing enterprises
  • explore

Thermal Conductivity by Guarded Heat Flow Meter (Single Point)


Test ID  
T-107

   
Description  
A sample measuring 2" in diameter is placed between the upper and lower plates of a guarded heat flow meter.  A thin layer of conductive grease is applied to the top and bottom of the sample and it is clamped in place.  The upper and lower plates are controlled at different temperatures, thus inducing heat flow through the sample.  A heated guard surrounding the sample reduces the lateral heat flow.  From the measure of heat flow, thermal conductivity is calculated.     
Standards  
ASTM E1530  
Data see sample data  
Solid thermal conductivity is reported. This data appears in Matereality as Thermal Conductivity.  
Replicates (# of times this test is repeated)  
1  
Samples  
2 - discs (50mm diam.) [details]

   

Application
 
This test is used to measure thermal conductivity of solid materials that range in thickness from 0.5 mm to 20 mm.