Charpy Pendulum Impact
Test ID
M-427
Description
A Dynatup POE 2000 apparatus is used to impact a specimen. Specimens are bars with dimensions specified in ASTM and notched accordingly. Sample is placed horizontally, with notch opposite of impact side. At set velocity and energy, pendulum is released to impact specimen. Test is complete when pendulum completes one swing. Five replicates are performed.
Options
Standards
- ASTM D6110
- ISO 179
Data
Break Type, Impact Strength. This data appears in Matereality as Charpy Impact.
Replicates
Number of times this test is repeated: 5
Samples
10 notched Charpy specimens [details]
Application
This test is used as a means of quality control or materials specification. It can, in some cases, be used on a finished product to determine if the part will fracture under the set conditions. Data gathered from this test are valuable for comparative purposes, to aid the engineer in choosing an appropriate material for a specific application.