Low Speed Multiaxial Punch


Test ID

M-411

Description

A multiaxial punch test is performed on a sample at a constant displacement rate. This rate is typically 5 mm/min. Force and displacement are recorded. The clamps and tups available for this test conform to ASTM D3763 and ISO 6603. Please specify which of these test standards should be followed. Please note that this is not a falling dart test and is instead performed at a forced velocity.

Options

Tests may be performed at different temperatures and displacement rates (up to 1 m/s)

Standards

  • ASTM D3763
  • ISO 6603-2

Data

Deflection at Maximum Load, Displacement v. Time Plot, Energy and Load v. Deflection Curves, Energy and Load v. Time Curves, Energy at Maximum Load, Energy v. Deflection Curves, Energy v. Time Curves, Failure Type, Load v. Displacement Plot, Maximum Load, Total Energy. This data appears in Matereality as Instrumented Dart Impact.

Replicates

Number of times this test is repeated: 5

Samples

10 impact plaques [details]

Application

The data from this test can be used to validate material models.